DMMCheck Plus

2015/05/20 By staze

DMMCheck PlusAs some may have realized, I have a decent number ((10, by last count)) of multimeters now but since all of them have been purchased second hand (eBay, or Goodwill), and several of them have been repaired, I had no real solid confidence in any of them as being highly accurate. Though, to be fair, I didn’t have any real reason to doubt them either. I did check in to see what calibration of one of my Keithley 196’s would cost, but the number was much more than I’d want to pay (I believe it was about $450). So, I’ve been eyeing a product called the DMMCheck Plus for a while. It provides a 5V DC and AC, 1mA DC and AC, some precision resistors (100R, 1k, 10k, 100k), and a 100Hz frequency to check your meters against. All of these are calibrated by the guy that makes them with his 8.5digit HP meter, that’s NIST traceable. Meaning I should be able to check against it for how far out of calibration any of my gear is.

I happened to have a PayPal balance, so I purchased my DMMCheck Plus through eBay and since they’re made up in southern Washington, I received it the day after shipped (I think from order to receipt was 3 days). The unit comes with a battery (which is how it’s calibrated), a hand filled in calibration sheet (showing the values for each test), number of burn in hours (mine was burned it for just over a month), and instructions. It’s really quite easy to use, and well worth the money, I think, to at least get a sense of how accurate your meters really are.

What I didn’t expect was what would be the most “in cal”. My Fluke 8060a that I purchased for about $40, replaced the caps on, and mainly have for nostalgia, was the most calibrated of my handheld meters. My Keithley 199 was the most in calibration of my bench meters. And amazingly (though not surprising given everything else I’ve seen with it), my Racal-Dana 1992 was amazingly accurate. The calibration of the 100Hz source says it should be 100.004608Hz, and my 1992 on a 10s gate showed 100.00463493Hz, or off by only 0.000026Hz ((Accounting for significant digits)). My other meters were all close enough for my work, and certainly not so far out that I’d pay for a calibration. I’ll definitely be double checking with this device now and again. Really makes me wish there was some ability to make a pocketable voltage reference that did 300mV, 3V, 30V, and 300V, as those are the ranges for my Keithley bench meters. But, battery driving for 30V and 300V, and getting a stable reference value might be tough. =(

Really, I’d highly suggest picking up one of these if you actually worry about your calibration for meters, or just have no idea how good or bad they are.

[xrr rating=5/5]

Filed Under: Reviews Tagged With: DMM, DMMCheck Plus, Fluke 189, Fluke 8060a, Fluke 87, Keithley, Racal-Dana 1992

Quick repair of Fluke 87 (Faded Digits)

2015/01/09 By staze

Fluke 87 Faded DigitsAt my suggestion, a coworker purchased a Fluke 87 (Series 1) off eBay, with the faded digit problem, for $70 a week ago. A very good deal. I assured him I could repair it.

The meter arrived, and really, there was very little fading, but it still needed “repair”. Rather than bring it home, I pulled it apart here and showed him what to do (should it happen again). Mr. Modemhead has a great article on this, but also there is a video from Fluke themselves on the issue, even though they’re incorrect on some facts with regards to the original 87 (they say there’s only one zebra strip, when there are two). Anyway, I’ve done the modemhead way, but I thought I’d try Fluke’s suggestion, and just clean the PCB side of the connection. I used some cotton swabs and 99% IPA, and after a couple minutes of scrubbing until clean, I put everything back together, and bingo! It works great. =)

Interestingly, this meter also had a slight issue with the buzzer/beeper. Turns out the speaker (Piezo) uses two rubber “nipples” to make contact with contact points on the back of the PCB. One of these nipples had been squashed/distorted over time and if you screwed the back of the meter on fully/snuggly, the nipple would distort enough to lose contact with the PCB. Simple solution was to leave the middle back screw loose, and place a piece of tape over the screw. Problem solved. =)

Filed Under: Electronics Tagged With: Faded Digits, Fluke 87, Multimeter, repair

2nd Fluke 87 True RMS Multimeter Repair

2014/06/21 By staze

Fluke 87Just posting a quite note that I recently purchased another Fluke 87 from eBay, this time at a more rational price of $75, with a fading display. With the wonderful help of Mr. Modemhead’s detailed explanation, I was able to get the screen back to factory quality, and the unit looks brand new!

Also cleaned up the yellow rubber boot with another page he published here. Meter looks brand new, took me maybe 25 minutes to do the screen repair. I’m pretty darn happy. I now have a Fluke 8060a, 87, and 189… I’m not sure they’re all considered the same family, but that’s how I’m going to look at it. =)

Filed Under: Electronics Tagged With: Fluke 87, Multimeter, repair

Weather

Categories / Archives

  • Apple
  • Coding
  • Electronics
  • Energy
  • Home Ownership
  • Miscellany
  • Politics
  • Prius
  • Sys Admin
  • Travel
  • Uncategorized
  • Work
  • June 2026
  • April 2026
  • August 2025
  • April 2025
  • January 2024
  • February 2021
  • July 2020
  • January 2020
  • April 2019
  • March 2018
  • February 2018
  • June 2017

Copyright © 2026 · Staze On Genesis Framework · WordPress · Log in